4002516 Autoclaves for sterilization
FEATURES
Stainless steel exterior and top.
Stainless steel lid. Lid, six closure point system.
Pulsed vacuum system and atmosphere purge.
External steam generator (except 150 litres autoclave).
Water reservoir for clean water for the steam generator boiler.
External water connection if required. 6 function modes.
RS232 for PC parameters printing (back side). Printer output.
USB output for PEN-DRIVEdata logging.
Connection for a second temperature sensor.
10 customer set programs.
Lateral external probe port (except 150 litres autoclave).
SAFETY Safety:
The lid cannot be opened if the chamber is under pressure.
Safety valve: Prevents over pressure over the maximum limit.
Safety pressure switch: switches the heater power off to if over pressure.
Safety Lid: insulated lid protection.
BACK VIEW
1. Mains input.
2. External water supply.
3. Safety valve.
4. Vacuum vapour valve.
5. Air in filter valve.
6. Safety thermostat.
7. Safety thermostat.
8. Empty tank valve.
9. Drain tank valve.
10. RS232 connector.
11. USB connector.
MICROPROCESSOR FEATURES
The microprocessorcontrols theperformingof theairextractionprocess, sterilizinganddrying,executing the selected procedures and commands by the user, adjusting the following parameters:
• Sterilization temperature from 105degree up to 134degree.
•Sterilization time from 3’ to 59’.
• Drying time from 20’ to 60’.
• Up to 10 user-configurable programs.
•Six operating modes: - Solid 1 - Solid 2 fractions. - Solid 3 fractions - Drying - Liquids - Liquids 2 probes
• Verification test - Vacuum - Bowie Dick.
•Settings: - Units-languages (Spanish,English, French). - Printer module On/Off. - Clock: date/time.
-External water: on/off.
• Password for configuration.
• Calibration (adjustment offset) - Temperature probe 1 - Temperature probe 2 - Water dosing. - Pressure
• Audible and visual alarm for overtemperature and insufficient temperature.
• Network failures detection and memorization
•Safety therrmostat by software
• Sterilization cycles data logging with capacity for over 500 cycles.
• Internal cleaning program
• Graphical representation of real time process
Error messages of functions and program execution are displayed and an acoustic alarm sounds. The vapour generator automatically cuts out.
MODELS:AUTESTER ST DRY PV lll 50
Part No.:4002516
Dry:Yes
Capacity litres:50
Ø Height(usable cm) :30 63
Height/Width/Depth (exterior cm) :117 48 62
Temp ° C:105 to 134
Water Reservoir H2O capacity.:10
Basket size:3
Power W:3000
Weight Kg: 108